34828 HYDRANALTMWater Standard 1.0
34847 HYDRANALTMWater Standard 0.1
34446 HYDRANALTMWater Standard 0.1 PC
These standards are used for testing the reliability of KF analyses according to different norms(e.g.ISO9000 or ISO17025)They are also suitable for water addition when carrying out a suitabilitytest accordingtoPh.Eur
Use
The ampoule has a predetermined breaking zone. Protect yourself by wearing gloves.Open the ampoule directly before useAtmospheric moisture can penetrate in anopen ampouleor very dry air can absorb water from theliquidand therefore falsify the results.For expiration data for the closed ampoule see label and Report of Analysis.
1. Snap off the top tag of the ampoule at the predetermined breaking zone.
2. Rinse a glass syringe with approx.0.5 mL standard
3. Draw the rest of the standard immediately into the syringe.
4. Carryout 2-3 control determinations.
For monitoring the system or after the anaylsis of a sample, add a known amount of Hydranal Water Standard and analyze it the same way as the sample. The amount of added water should be similar to the water amount in the sample.
The recovery rate should ideally be within the expanded measurement uncertainty stated on the Report of Analysis.Literature gives varying recommendations forrecoveryrateseg.the EuropeanPharmacopoeia:
For 1000μg water max.±25μg
For 100 ug water max.±10μg
Report of Analysis
Analyzed for Honeywell PMT Advanced Materials E&C,Offering Management,115 Tabor Road,Morris Plains, NJ 07950, USA
Product: HYDRANALM Water Standard1.0 Cat. No:34828 Lot:L2800
water content:1.003 mg/g expanded uncertainty:0.012 mg/g (k=2; 95% confidenceinterval)
determined by direct coulometric titration(Test Method 02 HYDRANAL version 13Nov.2017)
The water content is analyzed under ISO/EC 17025 accreditation by coulometric titration on 11 samples and is directly traceable to SI unit. Each sample is measured twice using two different coulometric instruments.The volumetric KF titration was used as a second technigue on a selected number of samples to confirm the reliability of the performance Both methods are verified against the external standard NIST SRM 2890.
Honeywell Specialty Chemicals Seelze GmbH Thomas WendtSupervisor Hydranal CoE Wunstorfer Str.40D-30926 Seelze