QCO-135 Low Level PAHs
Hazard: Irritant, Flammable
Solvent: Acetonitrile
Analyte |
Study Mean
μg/L
|
Certified Concentration
μg/L
|
Acceptance Limits
μg/L
|
1-Methylnaphthalene
|
0.00
|
0.00 +/- 0
|
0.00-0.00 |
2-Methylnaphthalene
|
0.00 |
0.00 +/- 0 |
0.00-0.00 |
Acenaphthene
|
9.41
|
12.0 +/- 0.112
|
4.03-14.5
|
Acenaphthylene
|
7.82
|
9.89 +/- 0.021
|
4.24-11.4
|
Anthracene
|
1.30
|
1.88 +/- 0.0175
|
0.551-2.55
|
Benzo(a)anthracene
|
1.18
|
1.47 +/- 0.0137
|
0.892-1.71
|
Benzo(a)pyrene
|
1.51
|
1.47 +/- 0.0137
|
0.627-1.81
|
Benzo(b)fluaranthene
|
1.16
|
1.01 +/- 0.0094
|
0.488-1.28
|
Benzo(g,h,i)perylene
|
1.26
|
1.08 +/- 0.0101
|
0.345-1.46
|
Benzo(k)fluoranthene
|
2.38
|
3.79 +/- 0.0353
|
1.93-4.83
|
Chrysene
|
0.435
|
0.500 +/- 0.0047
|
0.230-0.685
|
Dibenzo(a,h)anthrscene
|
2.66
|
2.58 +/- 0.0240
|
0.884-3.33
|
Fluoranthene
|
0389
|
1.20 +/- 0.0112
|
0.627-1.47
|
Fluorene
|
7.45
|
9.59 +/- 0.0833
|
4.08-11.1
|
lndeno(1,2,3-c,d)pyrene
|
3.37
|
3.87 +/- 0.0360
|
1.64-4.85
|
Naphthalene
|
5.15
|
7.36 +/- 0.0685
|
2.45-8.46
|
Phenanthrene
|
1.53
|
1.93 +/- 0.0180
|
0.962-2.31
|
Pyrene
|
2.90
|
3.38 +/- 0.0315
|
1.84-3.97
|
This CRM was manufactured by NSI Lab Solutions following quality procedures meeting the requirements of ISO 9001, ISO 17025, and ISO 17034. Acceptance limits are set at current industry standards. The study mean is set at the mean of an interlaboratory proficiency testing study with outlier rejection. This CRM is intended to be used to validate analytical methods, for detection limit studies, and far analyst proficiency testing. Certified concentration is the prepared concentration traceable to NIST.
Storage & Instructions For Use
Required storage is 2°C to 8aC.
Fill a 1000 mL Ciass A volumetric flask with 950 ml of organic free reagent water.
Allow the ampule concentrate to equilibrate to room temperature.
Open the ampule and pipet exactly 1.0 mL of the concentrate into the flask with a Class A pipet. Bring the flask to volume with reagent water.
Stopper the flask and mix well
This represents the sample for analysis.
Traceability Information
Analyte Source Materials: The highest purity analyte source materials are used in the manufacture of this CRM. Analyte source material purity and associated uncertainty has been analyticsIty verified against appropriate NIST SRMs, where available.
Balance: All analytical balances are calibrated on a semiannual basis by an ISO 17025 accredited calibration laboratory and are traceable to NIST. Traceable Calibratiort Certificate available upon request.
All balances are checked daily by an in-house standard operating procedure. The weights used for this daily verification are calibrated annually by an ISO 17025 accredited calibration laboratory and are certified traceable to NIST. Certificate of Calibration and Traceability available upon request.
Thermometer: Ail thermometers are NIST traceable through thermometers that are calibrated annually by an ISO 17025 accredited calibration laboratory.
Glassware: All glassware used in the manufacture of our samples is Class A. An in-house standard operating procedure is used to verify all glassware prior to it being placed into service. Volumetric pipetors are calibrated every four months by an ISO 17025 accredited calibration laboratory.
Homogeneity/Stability/Expiration
This CRM was thoroughly mixed in production. Batch homogeneity was established through analyses of samples chosen at random consistent with guide ISO-35:2017. The stability of this CRM is based on short-term and longterm monitoring of the certified concentration. The expiration date is guaranteed to be valid from the manufacture date and is based on results of long-term monitoring.
注:证书信息仅供参考,以产品附带证书为准。