gamma-HCH
Formula:C6H6Cl6
Mol. Welght:290.81
CAS No.:58-89-9
Storage Temperature:20℃±4℃
Storageandhandling:
The RM should be stored in the original sealed bottle at the temperature given above.After use the bottle should be tightly closed and protected from moisture and light .The experydate is valid for original sealed bottles under recommended storage conditions only.
Purity:98.87% (g/g)
Expanded Uncertainty U= 0.71% (g/g)
The uncertainty of this standard is calculated in accordance with the lSO Guide 34 and EURACHEM/CITAC Guide - Quantifying Uncertainty in Analytical Measurement, Second Edition.The expanded uncertainty is U(exp)=u(RM)x k, where k is the coverage factor at the 95% confidence level (k=2). Uncertainty u(RM)is based on the combination of the uncertainties associated with each indvidual operation involved in the analysis of the product u(RM)=vu(char)2+u(bb)2+u(lts)2+u(sts)2; u(char) is the uncertainty of charactersation; u(bb) uncertainty of homogeneity test;u(lts) uncertaintyof stability test long-term; u(sts) uncertainty of stablitytest short-term.u(lts) and u(sts) are not incuded in the calculation as the stablity statement is based on real evidence opposed to simulation.
Minimum sample: 1 mg is recommended as the minimal sample amount.If less material is used it is recommended to increase the ceritfied uncertainty by a factor of two for half sample and a factor of four for a quarter of sample.
Intended use: Use this RM as calibrant for chromatography or any other analytical technique.
Analytical Data
Traceabity of dhromatography: To the intemational System of Units (SI).
Instrumert:GC/FID
Detectionc:FID
Column:Optima-5MS,0.25 um, 0.25 mm
lnj.-VolL:1μl
Flow:1.0 mL/min
Ret.Time:11.54min
lnjector:320°C
Initial Temp:120ºCfor 4 min
End Temp:320°C for3 min
Gradient:15°C/min
Comment
Traceabity: The balances used are calibrated with weights traceable to the national standards (DKD).
Calibnated dass A glassware is used for volumetric measurements.
Water Content:<0.10% (g/g) by Karl-Fischer-Titration (U(exp) = 0.03% (g/g)).
ldentity:EA, NMR RT, IR
注:证书信息仅供参考,以产品附带证书为准。