PRODUCT DESCRIPTION
Product Code: Single Analyte lon Chromatography Solution
Catalog Number: ICCLO31
Lot Number: K2-CLOX653412
Matrix: H2O
Value l Analyte(s): 1 000 ug/mL
ea:Chlorate
Starting Material: Potassium chlorate
Starting Material Lot#t: 02407TF/ 2120
Starting Material Purity: 99.0000%
CERTIFIED VALUES AND UNCERTAINTIES
Certified Value: 1007± 6μg/mL
Certified Density: 0.999 g/mL (measured at 20 ± 1 ºC)
Assay Information:
Assay Method #1 1 007± 5 ug/mL
ICP Assay NIST SRM 3182 Lot Number: 060925
- The Calculated Value is a value calculated from the weight of a starting material that has been certifieddirectly vs. a National Institute of Standards and Technology (NIST) SRM/RM.See Sec 4.2 for balancetraceability.
TRACEABILITY TO NIST
- This product is traceable to NIST via an unbroken chain of comparisons. The uncertainties for each certifiedvalue are reported, taking into account the SRM/RM uncertainty error and the measurement, weighing andvolume dilution errors. In rare cases where no NIST SRM/RM are available, the term 'in-house std.' isspecified.
1.Thermometer Calibration
-All thermometers are NIST traceable through thermometers that are calibrated by an accredited calibration laboratory.
2.Balance Calibration
-All analytical balances are calibrated by an accredited calibration laboratory and procedure. The weights used for testing are annually compared to master weights and are traceable to NIST.
3.Glassware Calibration
-An in-house procedure is used to calibrate all Class A glassware used in the manufacturing and quality control of CRM/RMs
CHROMATOGRAM
INTENDED USE
For the calibration of analytical instruments and validation of analytical methods as appropriate
NSTRUCTIONS FOR THE CORRECT USE OF THIS REFERENCE MATERIAL
1.Storage and Handling Recommendations
-Store between approximately 4° - 30° C while in sealed TCT bag.
-While stored in the sealed TCT bag, transpiration of this CRM/RM is negligible. After opening the sealed TCT bag transpiration of the CRM/RM will occur, resulting in a gradual increase in the analyte concentration(s). It is the responsibility of the user to account for this effect. When the bottle is weighed both before and after being placed in storage, the mass difference observed will be a measure of transpiration mass loss.
-After opening the sealed TCT bag keep cap tightly sealed when not in use. Store and use at 20° ± 4° C. Do not pippette from the container. Do not return removed aliquots to container.
HAZARDOUS INFORMATION
-Please refer to the Safety Data Sheet for information regarding this CRM/RM.
HOMOGENEITY
-This solution was mixed according to an in-house procedure and is guaranteed to be homogeneous. Homogeneity data indicate that the end user should take a minimum sample size of 0.2 mL to assure homogeneity.
QUALITY STANDARD DOCUMENTATION
1.10CFR50 Appendix B - Nuclear Regulatory Commission
- Domestic Licensing of Production and Utilization Facilities
2.10CFR21 - Nuclear Regulatory Commission
-Reporting defects and Non-Compliance
3.ISO 9001 Quality Management System Registration
-QSR Certificate Number QSR-1034
4.ISO/IEC17025"GeneralRequirementsfortheCompetenceofTestingandCalibrationLaboratories"
Chemical Testing - Accredited / A2LA Certificate Number883.01
5.ISO Guide 34 "General Requirements for the Competence of Reference MaterialProducers"
Reference Material Producer - Accredited / A2LA Certificate Number883.02
CERTIFICATION, LOT EXPIRATION AND PERIOD OF VALIDITY
注:证书信息仅供参考,以产品附带证书为准。